
282
syngo Argus
• The Argus result table also shows normal ranges.
The range values are based on the study, Normal
human right and left ventricular mass, systolic func-
tion, and gender differences by cine magnetic reso-
nance imaging (see: Lorenz CH, Walker ES et al,
J Cardiovasc Magn Reson. 1999;1(1):7 – 21). This
study was performed on a group of 75 healthy, pre-
dominantly Caucasian subjects.
• If original axial CT images are loaded, Argus shows
both definitions, the PCS slice position and the CT
slice position. Argus uses the PCS slice position in its
messages.
• syngo Argus requires CT MPR images reconstructed
with any version from VA70 up to now. CT images
from other vendors, or CT images that have been
reconstructed with older versions do not contain the
cardiac phase in their DICOM parameter set. In this
case Argus will try to read the cardiac phase from the
image comment line. Therefore the image comment
line has to be modified manually.
• The format of the image comment line must be
"XXXX, ZZZ %" if the phase is specified in % of RR
interval "XXXX, ZZZ ms" if the phase is specified in
milliseconds.
• XXXX is alphanumeric text of arbitrary length with-
out comma signs, ZZZ denotes the value.
Example 1: this is just a text, 80%
Example 2: 60 Bpm, 233 ms
Commenti su questo manuale