Siemens SOMATOM Sensation Cardiac Version A60 Istruzioni per gli Interventi Pagina 281

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 600
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 280
syngo Argus
281
Hints in General
MPR series for Argus Ventricular Function analysis
should not contain overview images. It is recom-
mended, to deselect Include the reference image
when storing the range in the 3D configuration.
When series for Argus VF contain overview images,
it can happen that these images are sorted before
the ED/ES labeled images in the Argus splash matrix.
When computing the volume results then, Argus will
issue a warning "Volumes can not be computed for
non parallel slices. Not all results are available".
The results of the analysis are not affected.
syngo Argus uses a special image text configuration
which can not be changed by you and which can not
be filmed.
In reformatted images (MPR), Argus shows a slice
position in the image text. This slice position is given
in the patient coordinate system (PCS) as defined by
DICOM. The PCS slice position begins always with a
letter (A: anterior, P: posterior, F: feet, H: head, R:
right, L: left) followed by the coordinate e.g.
H 332.5.
If original axial CT images are loaded, Argus shows
both definitions, the PCS slice position and the CT
slice position. Argus uses the PCS slice position in its
messages.
Vedere la pagina 280
1 2 ... 276 277 278 279 280 281 282 283 284 285 286 ... 599 600

Commenti su questo manuale

Nessun commento