syngo Argus
281
Hints in General
• MPR series for Argus Ventricular Function analysis
should not contain overview images. It is recom-
mended, to deselect Include the reference image
when storing the range in the 3D configuration.
• When series for Argus VF contain overview images,
it can happen that these images are sorted before
the ED/ES labeled images in the Argus splash matrix.
When computing the volume results then, Argus will
issue a warning "Volumes can not be computed for
non parallel slices. Not all results are available".
The results of the analysis are not affected.
• syngo Argus uses a special image text configuration
which can not be changed by you and which can not
be filmed.
• In reformatted images (MPR), Argus shows a slice
position in the image text. This slice position is given
in the patient coordinate system (PCS) as defined by
DICOM. The PCS slice position begins always with a
letter (A: anterior, P: posterior, F: feet, H: head, R:
right, L: left) followed by the coordinate e.g.
H 332.5.
• If original axial CT images are loaded, Argus shows
both definitions, the PCS slice position and the CT
slice position. Argus uses the PCS slice position in its
messages.