HeartView CT
Cardiac Reconstruction
Axial Images
Reconstruction of axial images during the examination
are used for filming/archiving and for post-processing.
A. Filming and Archiving
Use a thicker slice thickness and recon increment for
filming and archiving of the images, e.g. 3 mm slice
thickness and 3 mm increment, to produce less
images.
You can send this recon job via the Auto Tasking tab-
card directly to the filmsheet or archiving node.
B. Calcium Scoring Postprocessing
Use a thicker slice thickness and a smaller recon incre-
ment to load the images afterwards in the application
syngo Calcium Scoring, e.g. 3 mm slice thickness and
1.5 mm increment.
C. 3D Postprocessing
Use a thinner slice thickness and recon increment for
post-processing of the images, e.g. 1.0 mm slice thick-
ness and 0.7 mm increment. These images are used for
post-processing on the 3D card or with syngo
InSpace4D.
You can select different image types, e.g. MIP or VRT to
get the best view of the coronary arteries.