UM344-2 ON-LINE AND OFF-LINE OPERATION
August 1995
5
-
7
SHORT ADDRESS 00
ENTER ADDRESS
EN-
< QUIT TER
NOTE
Only numbers may be used in the serial number
field.
Press SEL/CONT to display the Short Address screen.
Pressing SEL/END will end configuration of the transmitter
ID Block.
6. Enter or edit the transmitter short address using the EDIT,
QUIT, and ENTER keys as described below.
EDIT - display the Short Address edit screen. Use the < key
to edit by erasing one character in the address at a time,
starting with the last character. Type the new short address
on the keypad.
QUIT - return to the previous screen without saving the
address.
ENTER - store the new short address in the current MXC
On-Line Memory. The MXC will now show the Function
Block Menu.
5.1.1.4 Output Block
1. At the Function Block Menu, press NEXT to configure the
Output block and display the Failsafe Level screen.
2. Edit or change this value by scrolling through the selections
using the LAST OPTN and NEXT OPTN keys.
3. Press either SEL/END or SEL/CONT to return to the
Function Block Menu.
5.1.1.5 Alarm Block
There are two absolute alarms which may be configured as HI or
LO. The setpoint for the alarms is in the same units selected in
the operator display block.
1. Press SEL (F4) at the Alarm Block to display the Alarm 1
screen.
2. Press NEXT OPTN (F2) to enable or disable Alarm 1.
3. Press SEL/CONT (F4) to continue to the setpoint screen.
SHORT ADDRESS 00
SEL/ SEL/
EDIT END CONT
FAILSAFE LEVEL :
LOW
LAST NEXT SEL/ SEL/
OPTN OPTN END CONT
ALARM1 : ENABLE
LAST NEXT
SEL/
SEL/
OPTN OPTN
CONT
END
ALARM1 SETPOINT:
25.000 DEG F
SEL/
SEL/
CONTEND
EDIT