11.9 HART communication
HART communication
Load for a
HART communicator connection 230 … 1100 Ω
HART modem 230 … 500 Ω
Cable 2-wire, shielded: ≤ 3.0 km (1.86 miles), multi-
wired, shielded:
≤ 1.5 km (0.93 miles)
Protocol HART Version 5.x
PC/laptop requirements IBM-compatible, user memory > 32 MB, hard disk
>
70 MB, depending on the type of modem: RS 232
interface or USB connection, VGA graphic
Software for computer SIMATIC PDM
Technical data
11.9 HART communication
SITRANS P, DS III series with HART
194 Operating Instructions, 09/2012, A5E00047092-08